PIPETTE / DISPENSER / CUSTOM MADE & OEM

Select a Region / Language

Asia-Pacific

Europe, Africa, Mid East

Americas

Autoclavable Multi Channel Pipette | Micropipette

Nichipet EXⅡ MULTI

  • Autoclavable
  • UV Resistance
Dream Bank Point
8ch: 2pt / 12ch: 3pt

This product was discontinued at the end of September 2023.
Compatible tips, tip racks, stands, spare parts,
and consumables will continue to be available until September 2028.

Specifications

Cat. No Channel Volume Setting Volume range(μL) Increments(μL) Usable tips
00-NPM-8V 8 Digital Setting 0.5~10 0.01 BMT2-UT/UTWR、FUT/FUTRB
00-NPM-8S 5~100 0.1 BMT2-SG/SGR、FSG/FSGRB *
00-NPM-8L 20~200 0.2 BMT2-SG/SGR
00-NPM-8K 30~300 1 BMT2-K/KR
00-NPM-12V 12 0.5~10 0.01 BMT2-UT/UTWR、FUT/FUTRB
00-NPM-12S 5~100 0.1 BMT2-SG/SGR、FSG/FSGRB *
00-NPM-12L 20~200 0.2 BMT2-SG/SGR
00-NPM-12K 30~300 1 BMT2-K/KR
  • * The AC and CV values largely depend on the operator's and the environment under which the pipette is used.

AC & CV

Cat. No Volume(μL) Accuracy(%) Precision(%)
00-NPM-8V
00-NPM-12V
1 ±8.0 * ≦5.0 *
5 ±4.0 ≦2.0
10 ±2.0 ≦1.0
00-NPM-8S
00-NPM-12S
10 ±3.0 * ≦2.0 *
50 ±1.0 ≦0.8
100 ±0.8 ≦0.3
Cat. No Volume(μL) Accuracy(%) Precision(%)
00-NPM-8L
00-NPM-12L
20 ±3.0 ≦0.6
100 ±1.0 ≦0.4
200 ±0.9 ≦0.3
00-NPM-8K
00-NPM-12K
30 ±3.0 ≦1.0
150 ±1.0 ≦0.5
300 ±0.6 ≦0.3
  • *Each accuracy and precision are the values which come from measuring with our compatible tip.
  • *The AC and CV of NPM-8V and 12V below 1μL and NPM-8S and 12S below 10μL depend on the sampling skill and the sampling condition where pipettes are used.

Stands

00-MLT-STD2(ABS)
Cat. No Storing capacity
00-MLT-STD2 (ABS) 6pcs
* For only Multi-Channel,
3pcs is maximum storing capacity.

Tray

00-88-TY
Cat. No Q'ty
00-88-TY 50

Select a Region / Language

Asia-Pacific

Americas

Europe, Africa, Mid East