PIPETTE / DISPENSER / CUSTOM MADE & OEM

Select a Region / Language

Asia-Pacific

Europe, Africa, Mid East

Americas

Stroke Load Reduction | Autoclavable Multi Channel Pipette | Micropipette

Nichipet Air MULTI

  • Autoclavable
  • CE
Dream Bank Point
8ch: 3pt / 12ch: 3pt
Nichipet Air MULTI

Specifications

Cat. No channel Volume Setting Volume range(μL) Increments(μL) Usable tips Color
00-NAR-10M8 8 Digital Setting 0.5〜10 0.01 BMT2-UT / UTWR / UTWRB、
FUT / FUTRB
Pink
00-NAR-100M8 10〜100 0.1 BMT2-SG / SGR / SGRB、
FSE / FSERB
Orange
00-NAR-200M8 20〜200 0.2 BMT2-SG / SGR / SGRB、
FAG / FAGRB
Yellow
00-NAR-300M8 30〜300 1 BMT2-K/KR/KRB White
00-NAR-10M12 12 0.5〜10 0.01 BMT2-UT / UTWR / UTWRB、
FUT / FUTRB
Pink
00-NAR-100M12 10〜100 0.1 BMT2-SG / SGR / SGRB、
FSE / FSERB
Orange
00-NAR-200M12 20〜200 0.2 BMT2-SG / SGR / SGRB、
FAG / FAGRB
Yellow
00-NAR-300M12 30〜300 1 BMT2-K/KR/KRB White

AC & CV

Cat. No Volume (μL) Accuracy (%) Precision (%) Cat. No Volume (μL) Accuracy (%) Precision (%)
10μL
(00-NAR-10M8)
(00-NAR-10M12)
1 ±8.0 * ≦5.0 * 200μL
(00-NAR-200M8)
(00-NAR-200M12)
20 ±3.0 ≦1.5
5 ±4.0 ≦2.0 100 ±1.0 ≦0.7
10 ±2.0 ≦1.0 200 ±0.8 ≦0.3
100μL
(00-NAR-100M8)
(00-NAR-100M12)
10 ±3.0 ≦2.0 300μL
(00-NAR-300M8)
(00-NAR-300M12)
30 ±3.0 ≦1.0
50 ±1.0 ≦0.8 150 ±1.0 ≦0.5
100 ±0.8 ≦0.3 300 ±0.8 ≦0.3
  • * Each accuracy and precision are the values which come from measuring with our compatible tip.
  • * The AC and CV of NAR-10M8 and 10M12 below 1μL depend on the sampling skill and the sampling condition where pipettes are used.

Stands

Cat. No Storing capacity
00-NAR-STD 8pcs
  • Can be used for Nichipet Premium series and NICHIRYO Le.
  • Can be mounted on the stand with attaching a tip (only for Nichipet Air).

Tray

00-88-TY
Cat. No Q'ty
00-88-TY 50

Volume adjustment tool

Cat. No Q'ty
00-NAR-AT 1

Grease

Cat. No Q'ty
00-NAR-GRS 1